The Kanomax tracer gas diffuser kit is used in performing tracer gas tests in accordance with ANSI / ASHRAE Standard 110. The diffuser is designed to the specifications of standard drawing #110-83M.
The diffuser is placed in the fume hood and sulfur hexafluoride gas is injected at at supply pressure of 30 psig. The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be inserted.
Navigate the tab section below to see this product’s Features & Benefits, Specifications, Resources, and more.
Main Unit Specifications | |
---|---|
Critical Orifice | 4 Liters Per Minute |
Regulator | Dual Stage, Specialty Gas |
Flow Meter | Calibrated 150mm/200psi, Glass Tube |
Pressure Gauge | Large Dial, 0 to 60psi |
Warranty | 1 year |
Standard Accessories | Diffuser, Pressure Gauge, Shut-off Valve, In-line Flow Meter, 20 ft of Tubing, Tank Regulator for Gas Cylinder |
Our Tracer Gas Diffuser Kit, or Dif-Kit for short, is for use in performing the tracer gas test in accordance with ANSI/ASHRAE Standard 110-2016. The diffuser is made to the design and specifications of ASHRAE Standard Drawing #110-83M. Its purpose is to check fume hood performance to ensure that the hood is performing safely and to specifications.
Kanomax offers a complete fume hood testing solution. However, because not everyone needs to perform every aspect of the ASHRAE test we sell the kit as 5 separate modules so customers only need to buy the pieces required for their application. Please consult our sales specialists for the appropriate choice for your needs.
Anemomaster Professional: Our handheld anemometer is useful for performing a face velocity test at different points along the hood.
Fog Machine: This unit is used to perform a visualization test. Visible smoke is released inside the fume hood and monitored to see if any visible amount is escaping. It can also be used to show how the air is flowing inside the fume hood so trouble spots where airflow is insufficient can be detected.
Diffuser-kit: The core part of the DIF-KIT includes a flow meter, regulator and diffusion ejector. This portion of the kit is used to release the tracer gas inside the fume hood. Manikin-kit: This portion includes a test mannequin (to simulate a person standing at the fume hood), lab coat and mannequin stand.
Vapor Analyzer or Leakage detector: Instrument is calibrated to detect the tracer gas, such as SF6 (sulfur hexafluoride), and can be useful in determining exactly how much, if any, of the gas is escaping from the fume hood.
The flow meters that are part of the kit are also calibrated for use with the tracer gas.
Our sales representatives are here to help you find the right instrument for your application. Fast.
Request a quote today to find the product you need with expert assistance.