The 2700 series of Kanomax Gas Monitors can be configured with a wide range of single or multi-gas sensor heads. Each sensor head is simply plugged into the detector to begin monitoring. The detector does not require additional configuration allowing sensor heads to be swapped out easily in the field.
Increased sensitivity at a higher resolution is attained due to measurements being collected in ppb. Unit comes with a self calibration zero-out feature. The monitor is compatible with a full range of gas sensors. The optional remote sensor head adapter kit allows the sensor head to be located up to 46 ft. away.
Navigate the tab section below to see this product’s Features & Benefits, Specifications, Resources, and more.
Main Unit Specifications | |
---|---|
Measurement Units | PPM or mg/m3 |
Sensor Head | Removable and Replaceable |
Output | 0 to 5 V, Alarm output |
Data Logging | 8188 data points |
Power Supply | Li-Ion Battery or AC Adapter |
Dimensions (with sensor head) | 7.0 x 4.75 x 2.0 inches (195 x 122 x 54 mm) |
Weight (with sensor and battery) | 1 lbs (460 g) |
Standards | CE-mark Part 15 of FCC Rules EN 50082-1: 1997 EN 50081-1: 1992 |
Warranty | 1 year |
What's Included | Meter, Li-ion battery pack, Operation manual, AC adapter, PC connection cable, Data processing software |
Gas Monitor Sensor Heads | ||||||
---|---|---|---|---|---|---|
Sensor | Range (PPM) | Accuracy of Calibration | Min. Detection Limit (PPM) | Resolution (PPM) | Operational Temperature Range | Operational RH Range |
Ammonia | 0-100 | <±10% | 0.2 | 0.1 | 32-104°F | 10-90% |
Ammonia (leak) | 0-1000 | <±15% | 2 | 1 | 32-104°F | 10-90% |
Carbon Monoxide | 0-25 | <±0.5 ppm 0-5 ppm; | 0.05 | 0.01 | 32-104°F | 10-90% |
<±10% 5-25 ppm | ||||||
Carbon Monoxide | 0-100 | <±1 ppm 0-10 ppm; | 0.2 | 0.1 | 32-104°F | 10-90% |
<±10% 10-100 ppm | ||||||
Carbon Monoxide | 0-1000 | <±10 | 1 | 1 | 32-104°F | 10-90% |
Carbon Dioxide | 0-2000 | <±10 ppm +5% | 10 | 1 | 32-104°F | 10-90% |
Carbon Dioxide | 0-5000 | <±20 ppm +5% | 20 | 1 | 32-104°F | 10-90% |
Chlorine | 0-10 | <±10% | 0.02 | 0.01 | 32-104°F | 10-90% |
Formaldehyde | 0-10 | <±0.05 ppm 0-0.5 ppm; | 0.01 | 0.01 | 32-104°F | 10-90% |
<±10% 0.5-10 ppm | ||||||
Hydrogen | 0-5000 | <±10 | 5 | 1 | 32-104°F | 10-90% |
Hydrogen Sulfide | 0-10 | <±0.05 ppm 0-0.5 ppm; | 0.04 | 0.01 | 32-104°F | 10-90% |
<±10% 0.5-10 ppm | ||||||
Hydrogen Sulfide | 0-100 | <±0.5 ppm 0-5 ppm; | 0.04 | 0.01 | 32-104°F | 10-90% |
<±10% 5-100 ppm | ||||||
Methane | 0-10000 | <±15% | 10 | 1 | 32-104°F | 10-90% |
Nitrogen Dioxide | 0-1 | <±0.02 ppm 0-0.2 ppm; | 0.005 | 0.001 | 32-104°F | 10-90% |
<±10% 0.2-1 ppm | ||||||
Ozone | 0-0.15 | <±0.005 ppm | 0.001 | 0.001 | 32-104°F | 10-90% |
Ozone | 0-0.5 | <±0.008 ppm 0-0.1 ppm; | 0.001 | 0.001 | 32-104°F | 10-90% |
<±10% 0.1-0.5 ppm | ||||||
Ozone | 0-10 | <±0.1 ppm 0-1 ppm; | 0.02 | 0.01 | 32-104°F | 10-90% |
15% 1-10 ppm | ||||||
Perchloroethylene | 0-200 | <±5 ppm 0-50 ppm; | 1 | 1 | 32-104°F | 10-90% |
<±10% 50-200 ppm | ||||||
Sulphur Dioxide | 0-10 | <±0.05 ppm 0-0.5 ppm; | 0.04 | 0.01 | 32-104°F | 10-90% |
<±10% 0.5-10 ppm | ||||||
Sulphur Dioxide | 0-100 | <±0.5 ppm 0-5 ppm; | 0.4 | 0.1 | 32-104°F | 10-90% |
<±10% 5-100 ppm | ||||||
VOC (GSS sensor) | 0-25 | <±10% | 0.1 | 0.1 | 32-104°F | 10-90% |
VOC (GSS sensor) | 0-500 | <±10% | 1 | 1 | 32-104°F | 10-90% |
VOC (GSS sensor) | 0-20 | <±10% | 0.02 | 0.01 | 32-104°F | 10-90% |
VOC (GSS sensor) | 0-1000 | <±10% | 0.2 | 0.1 | 32-104°F | 10-90% |
Temperature | -20-100°C | <± 0.3°C | 0.01°C | -40-248°F | 0-105% | |
Humidity | 0 to 100% | <± 2% RH | 1% RH | -40-248°F | 0-105% |
Gas Monitor Laser Particle Sensor Head | ||||||
---|---|---|---|---|---|---|
Sensor Type | Range (mg/m3) | Accuracy of Calibration | Min. Detection Limit (mg/m3) | Resolution (mg/m3) | Operational Temperature Range | Operational RH Range |
LPC | 0.001-1.000 | ±0.005 mg/m3 +15% | 0.001 | 0.001 | 0-40°C | 0-90% |
Our portable gas detectors are designed for easy and affordable measurement of indoor and industrial air quality.
They are used by health and safety experts at industrial and commercial sites as well as researchers who are interested in measuring hot spots and testing personal exposure to harmful indoor gases.
Our sales representatives are here to help you find the right instrument for your application. Fast.
Request a quote today to find the product you need with expert assistance.